dc.identifier.citation | Pachao, A. (2016). Evaluación de desempeño de los sistemas de medición de análisis bioquímicos del laboratorio clínico Proyecta–Sucursal Cajamarca- para asegurar la calidad analítica de los resultados (Tesis de licenciatura). Repositorio de la Universidad Privada del Norte. Recuperado de https://hdl.handle.net/11537/10908 | es_PE |
dc.identifier.other | 670.42 PACH/E 2016 | es_PE |
dc.identifier.uri | https://hdl.handle.net/11537/10908 | |
dc.description.abstract | RESUMEN
La dirección de los laboratorios clínicos Proyecta DX conocedora de las desviaciones que
adolecía respecto al aseguramiento de la calidad de los resultados emitidos y siendo estos
resultados de vital importancia para las decisiones médicas, decidió enfocarse en la
solución de este problema.
Para solucionar estas desviaciones vio conveniente realizar una evaluación de desempeño
de los sistemas de medición iniciando con el método de glucosa (por ser el análisis con
mayor demanda); esta evaluación permitiría obtener información relevante para diseñar un
adecuado programa de control de calidad interno y finalmente lograr asegurar la calidad
analítica de sus resultados emitidos. Esta investigación requirió el uso de diferentes guías
y protocolos internacionales, tales como: User verification of performance for precission
and trueness del CLSI (Clinical Laboratory Standard Institute) y la guía para la validación y
la verificación de los procedimientos de examen cuantitativo empleados por el laboratorio
clìnico del EMA (entidad mexicana de acreditación). Para el análisis de los datos de sueros
de controles normales y patológicos se aplicaron herramientas gráficas y estadísticas.
Los resultados de esta evaluación comprobaron que el desempeño del sistema de
medición para el ensayo de glucosa cumplió satisfactoriamente los criterios especificados
por el fabricante respecto a precisión, veracidad, linealidad y el error total, además brindó
información necesaria para diseñar un programa de control de calidad interno el cual fue
utilizado para evaluar los datos de los meses de noviembre y diciembre del 2015 en el
laboratorio de Proyecta DX Sede Cajamarca, demostrando que estos fueron confiables y
asegurando la calidad analítica de los mismos ya que se pudo detectar la influencia de los
errores sistemáticos y aleatorios en los procesos de análisis.
Siendo esta evaluación de los sistemas de medición un punto neurálgico del aseguramiento
de la calidad de los resultados es necesario e imperativo continuar con su evaluación para
los demás métodos de ensayo en los laboratorios clínicos Proyecta en cada una de sus
Sedes. | es_PE |
dc.description.abstract | ABSTRACT
The management of clinical laboratories Proyecta DX aware of vitiating deviations regarding
quality assurance of the results being issued and the results of vital importance for medical
decisions, decided to focus on solving this problem.
To improve these desviations saw fit to make an assessment of performance measurement
systems starting with the method of glucose (being the higher demand analysis ) ; this
evaluation would obtain relevant information to design an appropriate program of internal
quality control and finally achieve ensure the analytical quality of the results issued . This
research required the use of different guidelines and protocols , such as : User verification
of performance for precission and trueness of CLSI (Clinical Laboratory Standard Institute)
and guidance for the validation and verification of procedures for quantitative examination
used by the laboratory clinical EMA (Mexican accreditation entity) . For data analysis of
serum from normal and pathological control graphical and statistical tools applied.
The results of this evaluation found that the performance measurement system for glucose
test successfully met the criteria specified by the manufacturer with respect to precision ,
accuracy, linearity and the total error also provided information needed to design a program
of quality control internal which was used to evaluate the data for the months of November
and December 2015 in the laboratory of Proyecta DX See Cajamarca, demonstrating that
these were reliable and ensuring the analytical quality of them since it could detect the
influence of errors systematic and random processes analysis.
As this assessment of measurement systems a focal point of quality assurance results is
necessary and imperative to continue their evaluation for other test methods in clinical
laboratories Proyecta in each of their headquarters. | es_PE |
dc.description.uri | Tesis | es_PE |
dc.format | application/msword | es_PE |
dc.format | application/pdf | es_PE |
dc.language.iso | spa | es_PE |
dc.publisher | Universidad Privada del Norte | es_PE |
dc.rights | info:eu-repo/semantics/openAccess | es_PE |
dc.rights.uri | https://creativecommons.org/licenses/by-sa/4.0/ | * |
dc.source | Universidad Privada del Norte | es_PE |
dc.source | Repositorio Institucional - UPN | es_PE |
dc.subject | Evaluación del desempeño | es_PE |
dc.subject | Laboratorios de química | es_PE |
dc.subject | Sistema de gestión de calidad | es_PE |
dc.title | Evaluación de desempeño de los sistemas de medición de análisis bioquímicos del laboratorio clínico Proyecta–Sucursal Cajamarca- para asegurar la calidad analítica de los resultados | es_PE |
dc.type | info:eu-repo/semantics/bachelorThesis | es_PE |
thesis.degree.grantor | Universidad Privada del Norte. Facultad de Ingeniería | es_PE |
thesis.degree.level | Título Profesional | es_PE |
thesis.degree.discipline | Ingeniería Industrial | es_PE |
thesis.degree.name | Ingeniero Industrial | es_PE |
dc.publisher.country | PE | es_PE |
dc.subject.ocde | https://purl.org/pe-repo/ocde/ford#2.11.04 | es_PE |
thesis.degree.program | Pregrado | es_PE |
dc.description.sede | Cajamarca | es_PE |
renati.advisor.dni | 40043738 | |
renati.author.dni | 25838856 | |
renati.discipline | 722026 | es_PE |
renati.juror | Sisniegas Noriega, Karla Rossemary | |
renati.juror | Mendoza Azañero, Ana Rosa | |
renati.juror | Rojas Gutierrez, Erika Liliana | |
renati.level | https://purl.org/pe-repo/renati/level#tituloProfesional | es_PE |
renati.type | https://purl.org/pe-repo/renati/type#tesis | es_PE |